Arrhythmia impact on ischemia detection by CMR

Impact of arrhythmia on diagnostic performance of adenosine stress CMR in patients with suspected or known coronary artery disease

Simon Greulich, Hannah Steubing, Stefan Birkmeier, Stefan Grün, Kerstin Bentz, Udo Sechtem and Heiko Mahrholdt

Journal of Cardiovascular Magnetic Resonance 2015, 17:94 

Full PDF here

Background

The diagnostic performance of adenosine stress cardiovascular magnetic resonance (CMR) in patients with arrhythmias presenting for work-up of suspected or known CAD is largely unknown, since most CMR studies currently available exclude arrhythmic patients from analysis fearing gating problems, or other artifacts will impair image quality. The primary aim of our study was to evaluate the diagnostic performance of adenosine stress CMR for detection of significant coronary stenosis in patients with arrhythmia presenting for 1) work-up of suspected coronary artery disease (CAD), or 2) work-up of ischemia in known CAD.

Methods

Patients with arrhythmia referred for work-up of suspected CAD or work-up of ischemia in known CAD undergoing adenosine stress CMR were included if they had coronary angiography within four weeks of CMR.

Results

One hundred fifty-nine patients were included (n = 64 atrial fibrillation, n = 87 frequent ventricular extrasystoles, n = 8 frequent supraventricular extrasystoles). Of these, n = 72 had suspected CAD, and n = 87 had known CAD. Diagnostic accuracy of the adenosine stress CMR for detection of significant CAD was 73 % for the entire population (sensitivity 72 %, specificity 76 %). Diagnostic accuracy was 75 % (sensitivity 80 %, specificity 74 %) in patients with suspected CAD, and 74 % (sensitivity 71 %, specificity 79 %) in the group with known CAD. For different types of arrhythmia, diagnostic accuracy of CMR was 70 % in the atrial fibrillation group, and 79 % in patients with ventricular extrasystoles. On a per coronary territory analysis, diagnostic accuracy of CMR was 77 % for stenosis of the left and 82 % for stenosis of the right coronary artery.

Conclusion

The present data demonstrates good diagnostic performance of adenosine stress CMR for detection of significant coronary stenosis in patients with arrhythmia presenting for work-up of suspected CAD, or work-up of ischemia in known CAD. This holds true for a per patient, as well as for a per coronary territory analysis.

Newsbeat Banner

SCMR Newsletter brought to you by NewsBeat

Read the current SCMR NewsBeat to stay up to date with the latest SCMR News!